Z-1496-2020: Artis Q.zen Biplane, Model No. 10848355 Artis Q.zen Ceiling, Model No. 10848354 Artis Q.zen Floor, Model

Recall record Z-1496-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1496-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Mar 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software issue could potentially cause the stand and table movements to be blocked.

Source facts

Action
On February 14, 2020, the firm distributed Urgent Medical Device Correction letters to customers. Customers were informed of the software error which affects the listed systems. In case of software failure, planned procedures may have to be terminated and performed on an alternative x-ray system. Customers were instructed to do the following: Issue # 1: Potential issue with the Message "Stand battery charging: DO NOT POWER OFF" on Artis zeego systems Customer Action: Please check the system messages before power off/shutdown of the system. If possible, do not switch off the system while the message "Stand battery charging: DO NOT POWER OFF" is displayed and wait until the battery has been charged or replaced. Issue # 2: Potential issue of blocked stand movements on Artis zeego systems Issue # 3: Potential issue of blocked table movements on Artis zee/ Q/ Q.zen systems with Siemens Healthineers tables For Issues #2-3, customers may continue to use their systems, but in case the problem occurs, customers are instructed to follow standard emergency procedures in place in the event of system failures. Customers are encouraged to ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. Siemens will correct the software error via a software update. Siemens' service organization will contact customers to arrange a date to perform this corrective action. For an earlier appointment, please contact the firm's service organization at 1-800-888-7436.
Initiated
2020-02-14
Root cause
Software design
Terminated
2021-04-20
Distribution
US Nationwide distribuiton.
Recall number
Z-1496-2020
Recall status
Terminated
CDRH record ID
180017
FDA classified
2020-03-13
Recall event number
85071

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗