Open original FDA record ↗Recall record
- FDA identifier
- Z-1496-2021
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Apr 26, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a malfunction of the belts which may result in limited functionality of the Artis system up to system failure. Unintended movement of the C-Arm may cause a hazardous situation to the patient, operator, or staff members and may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Source facts
- Action
- Siemens Medical Solutions USA, Inc. issued Urgent Medical Device Correction dated 3/303/21 (AX068/20/S). The letter states reason for recall, health risk and action to take: Use only recommended substances for cleaning and disinfection. Do not let cleaning liquids seep into the openings of the system, e.g., air openings, gaps between covers. Observe the attached cleaning and disinfection instructions. An addendum to the system Operator Manual explaining appropriate cleaning in detail has been released via Update Instruction AX068/20/S and is attached to this letter. This letter and the attached addendum should both be filed with the system documentation. Please feel free to contact our service organization at 1-800-888-7436 should you have any questions. The addendum will increase awareness concerning the cleaning process and mitigate the occurrence of the potential issue. There are no risks for patients who have previously been examined or treated. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein. Promptly notify and instruct accordingly all the staff at your organization who need to be aware of this notice and will comply with the recommendations therein. Please ensure that this safety advisory is retained in your product related records appropriately. In addition, if you may have further distributed this product, please identify your customers, and notify them at once of this product recall. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner. We would also request you to inform us of the identity of the device's new owner where possible. Please forward this safety information to any other organizations that could be affected by this measure.
- Initiated
- 2021-03-30
- Root cause
- Equipment maintenance
- Terminated
- 2022-07-12
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1496-2021
- Recall status
- Terminated
- CDRH record ID
- 186757
- FDA record created
- 2021-04-26
- Recall event number
- 87683