Z-1497-2019: Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine

Recall record Z-1497-2019 for Abbott Gmbh & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1497-2019
Company in FDA record
Abbott Gmbh & Co. KG
FDA record creation date
May 30, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.

Source facts

Action
Abbott notified customers on about 04/05/2019 via "Product Correction Urgent - Immediate Action Required" letter. The letter instructed customers to power off their Alinity i System if they experience unexpected reagent supply temperature errors, detect a burning smell or visible smoke. They were also instructed that their Abbott representative will contact them to schedule the replacement of the impacted parts, to forward the Product Correction letter if the device has been further distributed, retain the letter for laboratory records, and complete and return the attached Customer Reply form. Questions can be directed to Customer Service at at 18774ABBOTT (available 24 hours a day, 7 days a week) for US customers. For customers outside the US, they are instructed to contact their local area Customer Service.
Initiated
2019-04-05
Root cause
Under Investigation by firm
Terminated
2022-09-28
Distribution
Worldwide Distribution - US Nationwide to Oklahoma and Texas. Distributed worldwide to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, MADAGASCAR, MEXICO, NETHERLANDS, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UAE, UNITED KINGDOM, VIETNAM.
Recall number
Z-1497-2019
Recall status
Terminated
CDRH record ID
171991
FDA record created
2019-05-30
Recall event number
82595

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗