Z-1497-2021: Interventional Fluoroscopic X-Ray System ARTIS icono biplane- 11327600 ARTIS icono floor- 11327700 ARTIS

Recall record Z-1497-2021 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1497-2021
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Apr 26, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens has become aware of a potential issue with software version VE20C. Planned procedures may have to be terminated and performed on an alternative x-ray system.

Source facts

Action
On 3/19/2021, Siemens Healthineers, issued an "URGENT: MEDICAL DEVICE CORRECTION" notice via letter dated 3/192021 to customers informing them of a possible hazard to patients, operators, or other persons and equipment due to a software error which affects ARTIS icono /pheno systems with software VE20C. The customer are instructed to do the following: The user should be aware of the limited system functionality in backup mode and bypass fluoroscopy operating mode (see also Operator Manual, Chapter 6.7: System Operating Modes). In some rare cases, the automatic recovery phase may not be successful within 4 minutes. If necessary, it may be continued with manual recovery according to the Operator Manual instructions, Chapter 3.13: Recovering functions in case of system failures. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. Promptly notify and instruct accordingly all the staff at your organization who need to be aware of this notice and will comply with the recommendations therein. Please ensure that this safety advisory is retained in your product related records appropriately. In addition, if you may have further distributed this product, please identify your customers, and notify them at once of this product recall. If the device has been sold and is therefore no longer in your possession, please forward this notice to the new owner. Siemens will update the affected software to prevent this non-conformity from reoccurring via Update Instruction AX023/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. If you have any further questions, contact the Siemens Medical Solution USA, Inc., at 610-448-6478 or email: Meredith.adams@siemens-healthineers.com.
Initiated
2021-03-19
Root cause
Software design
Terminated
2021-11-09
Distribution
Worldwide distribution: US (nationwide) including states of: CA, CO, DE, FL, IL, IN, KY, MD, MI, NC, NE, NJ, NY, OH, TN, TX, UT, WI, and OUS (countries) of: AT, BE, ES, IT, FI, CH, FR, GB, SE, DE, DK, IE and NL.
Recall number
Z-1497-2021
Recall status
Terminated
CDRH record ID
186408
FDA record created
2021-04-26
Recall event number
87620

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗