Open original FDA record ↗Recall record
- FDA identifier
- Z-1497-2026
- Company in FDA record
- Ion Beam Applications S.A.
- FDA classification date
- Mar 6, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.
Source facts
- Action
- On February 3, 2026 Urgent Medical Device Correction letters were sent to customers. Letter provides awareness of the Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. On-site modification of the devices will be performed by IBA with a TSS Configuration update. The update will be deployed at the impacted sites by June 2026.
- Initiated
- 2026-02-03
- Root cause
- Software design
- Distribution
- Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
- Recall number
- Z-1497-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218471
- FDA classified
- 2026-03-06
- Recall event number
- 98364