Skip to content

Public FDA record

Z-1497-2026: IBA Proton Therapy System - PROTEUS 235

Recall record Z-1497-2026 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1497-2026
Company in FDA record
Ion Beam Applications S.A.
FDA classification date
Mar 6, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was identified that eh Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. In this condition, the TSS does not interrupt the beam delivery if the Healthy Signal from an external gating device is lost. This may lead to a risk of mistreatment when external gating devices are used.

Source facts

Action
On February 3, 2026 Urgent Medical Device Correction letters were sent to customers. Letter provides awareness of the Universal Beam Triggering Interface (UBTI) is disabled in the Therapy Safety System (TSS) configuration. On-site modification of the devices will be performed by IBA with a TSS Configuration update. The update will be deployed at the impacted sites by June 2026.
Initiated
2026-02-03
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the state of FL and the countries of Italy and Japan.
Recall number
Z-1497-2026
Recall status
Open, Classified
CDRH record ID
218471
FDA classified
2026-03-06
Recall event number
98364

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗