Z-1498-2020: LivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to circulate water

Recall record Z-1498-2020 for LivaNova USA Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1498-2020
Company in FDA record
LivaNova USA Inc.
FDA record creation date
Mar 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.

Source facts

Action
On 02/24/2020, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter to customers informing them that the firm is executing a voluntary correction for their Heater-Cooler Systems and providing customers with a recently cleared operating instructions that includes a validated instructions for cleaning and disinfecting the external surfaces, connectors, fittings, and water circuits. The firm is instructing customers to discard or destroy older versions of the operating instruction and advising that all Heater-Cooler Systems in use at their facility should be upgraded with the 3T Aerosol Collection Set (vacuum canister and internal seal) to further reduce the risk of potential aerosol emission form the 3T system. These corrections apply to all 3T Systems currently in distribution, which includes upgrading any loaners that were previously distributed or that are pending distribution. Actions to be Taken by the Customer/User Use the Validated Operating Instructions Provided with this Customer Letter " The FDA cleared Operating Instructions are found in Attachment 1 as well as at Product Resources on www.livanova.com. - Please read and follow the instructions carefully. - Section 2.1.5 Integrating the HC3T into your facility provides guidance on the use of the 3T System in your facility, highlighting critical tasks that users had the most difficulty with during human factors testing. - Section 6 Maintaining the heater-cooler provides guidance on the maintenance of the 3T System including the validated cleaning and disinfection instructions. - Section 7.1.4 General performance data describes performance testing results that emphasize the importance of following the cleaning and disinfection instructions and consistently using the 3T Aerosol Collection Set during surgeries. Specifically, these results describe how (1) the disinfection and water maintenance process maintains water quality and (2) the 3T Aerosol Collection Set design change reduces the
Initiated
2020-02-24
Root cause
Device Design
Terminated
2022-08-04
Distribution
U.S. Nationwide distribution including Puerto Rico. This matter was triggered by the clearance of K191402, thus only the United States, including Puerto Rico, is affected. O.U.S. - none
Recall number
Z-1498-2020
Recall status
Terminated
CDRH record ID
180090
FDA record created
2020-03-13
Recall event number
85098

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗