Z-1498-2022: Atellica CH Iron_2 (Iron_2) is for in vitro diagnostic use in the quantitative determination of iron in h

Recall record Z-1498-2022 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1498-2022
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Aug 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Falsely Elevated Atellica CH Microalbumin_2 (ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately after an iron test.

Source facts

Action
Siemens issued Urgent Medical Device Correction (UMDC) Letter by FedEx beginning on 06/28/2022. Letter states reason for recall, health risk and action to take: For customers operating with Atellica Software v.1.25.2 and lower, the resolution of this issue will be implemented in Atellica Software v1.25.3 which will be available soon. In the interim, please follow the instructions in the Additional Information section. Customers who are operating with Atellica Software v1.26 will receive further information when a software update to resolve the issue is available. For laboratories operating with Atellica Software v1.25.2 and below and Atellica Software v1.26 follow the workaround instructions in the Additional Information section until a future version of software is available. If your laboratory has multiple Atellica CH 930 Analyzers, Siemens recommends testing the Atellica CH ALB_2 assay on a separate analyzer from the Iron_2 assay. " If you choose not to separate the assays as indicated above, batch testing of Atellica CH ALB_2 may be considered. " If Iron_2 and ALB_2 will be processed on the same Atellica CH analyzer, an RPC2 wash mitigation must be initiated after processing Iron_2 and prior to processing ALB_2. Note: Any of the following will initiate the RPC2 wash: " After the Atellica CH 930 Analyzer has been in standby for 12 minutes. " After completion of any Open Channel assay. " Restarting the Atellica CH 930 Analyzer. Refer to the Atellica Solution Online Help for instructions on system restart Please review this letter with your Medical Director. Perform the instructions provided in Additional Information. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support
Initiated
2022-06-28
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of AR, AU, AT, BS, BH, BD, BE, BR, BG, BF, CA, CL, CO, HR, CW, CZ, DK, EC, EG, EE, FI, FR, DE, GR, GP, HK, HU, IN, IR, IQ, IE, IL, IT, JP, JO, KE, KW, LV, LY, LT, MY, MX, MA, NL, NZ, NO, OM, CN, PK, PY, PE, PH, PL, PT, QA, KR, RO, RU, SA, RS, SG, SK, ZA, ES, SE, CH, TW, TH, TR, AE, UG, GB, UY, VA, VN.
Recall number
Z-1498-2022
Recall status
Open, Classified
CDRH record ID
194780
FDA record created
2022-08-04
Recall event number
90559

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗