Skip to content

Public FDA record

Z-1498-2026: Raz-AT (Attendant Tilt) Mobile Shower Commode Chair, Catalog Number Z300 UDI-DI code: (01)00628758003007

Recall record Z-1498-2026 for Raz Design Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1498-2026
Company in FDA record
Raz Design Inc
FDA classification date
Mar 6, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a modification by a vendor and may be reinstalled without properly engaging the outside rear seat brackets that are used to keep the seat in place.

Source facts

Action
On 01/05/2025, the firm email an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them that an incident was reported to Raz involving a Raz Mobile Shower Commode Chair with a certain Front and Rear Open (FAR) Seat. This incident in Canada involved an unauthorized modification of a FAR seat by a vendor, followed by reinstallation without properly engaging the outside rear seat brackets that keep the seat in place. This is not a supported configuration, and it resulted in one side of the seat halves slipping off the seat frame, which caused an injury to the user. Customers are instructed to: 1 Notify impacted customers using this Correction Notice and request that they return the Acknowledgement and Receipt Form to you. 2 Raz will provide items to facilitate Correction. The items will be provided directly to the Sales Rep or to Dealer, whichever will execute the Correction action: 3 Sales Rep or Dealer shall conduct a service visit with the impacted end-user. For users/caregivers: 1 Be sure that the FAR seat is properly installed and latched before a user is transferred onto the chair. The rear seat attachment bracket must be fully hooked under the rear seat support tube(s). The front seat attachment brackets must be securely latched around the front seat support tubes. Both front and rear seat attachment brackets should be secured in place by tightening the 8 screws on the underside of the seat. 2 if seat needs to be removed for cleaning, follow instructions as per User & Dealer Manual proper removal and installation of the seat to ensure safe use of the chair prior to the installation of the FAR Seat Retrofit Kit. 3 If you do not have a User and Dealer Manual, download a copy from www.razdesigninc.com, or contact Sales Rep or Dealer. For questions or further assistance - contact Raz Design Inc by email at arahat@razdeigninc.com
Initiated
2026-01-05
Root cause
Use error
Distribution
U.S. Nationwide distribution in the states of CO, FL, IL, MA, MI, MN, MO, NC, NE, NH, NJ, NY, OR, PA, TX, WA, and WI.
Recall number
Z-1498-2026
Recall status
Open, Classified
CDRH record ID
218028
FDA classified
2026-03-06
Recall event number
98254

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.