Z-1499-2020: HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3

Recall record Z-1499-2020 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1499-2020
Company in FDA record
Philips North America, LLC
FDA record creation date
Mar 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical shock, even though the device did not have visible external damage or the Ready for Use ( RFU ) indicator on the unit did not immediately indicate a problem. One of these reports involved a death of a patient who could not be resuscitated.

Source facts

Action
On February 26, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that if the device is dropped or subjected to a severe mechanical shock, it may suffer internal damage even in the absence of visible external damage or the Ready for Use indicator unit does not immediately indicate a problem. Customers were instructed to inform all users that if the device is dropped or subjected to severe mechanical shock and the exterior case is still intact, they should immediately perform an operational check as described in the IFU section "Performing the Operational Check" in the Maintenance Chapter. The unit should be taken out of service and Philips Customer Service contacted if the unit is visibly damaged or if the device fails the operational check, i.e., if the RFU indicator changes to a red-X or the device emits a periodic audible chirp, as described in the IFU. Customers were instructed to insert a copy of the customer notice into each copy of the HeartStart MRx IFU. If you need further information or support concerning this notification, please contact your local Philips representative or call the firm at 1-800-722-9377.
Initiated
2020-02-26
Root cause
Device Design
Terminated
2022-12-08
Distribution
US Nationwide domestic distribution. Worldwide foreign distribution.
Recall number
Z-1499-2020
Recall status
Terminated
CDRH record ID
179524
FDA record created
2020-03-13
Recall event number
84841

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗