Skip to content

Public FDA record

Z-1499-2022: Craftmatic Model 1 Base for Dual Bed Configurations

Recall record Z-1499-2022 for Craftmatic Industries, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1499-2022
Company in FDA record
Craftmatic Industries, Inc.
FDA record creation date
Aug 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potential for dual-configured beds to separate or slip away from one another.

Source facts

Action
The firm initiated the recall by email on 03/24/2017. The notice explained the issues and provided the following actions: "1. If your bed or beds is/are on a smooth surface or on carpeting that does not secure the position of the bed(s), we require that rubber castor cups be laced under the legs of the bed. 2. If you purchased a dual system (two beds side by side with two individual mattresses) and your beds are not connected so as to prevent separation, we will assist you in connecting them with our recommended strapping. Craftmatic will make appropriate arrangements to ensure that your dual system beds are properly connected or secured without cost to you. 3. If you no longer are in possession of the bed(s) you received form Craftmatic it is IMPORTANT that you provide Craftmatic with the contact information for the current user of the bed or beds. Likewise, if the bed has been discarded and is not available for use by anyone it is IMPORTANT that you inform Craftmatic. 4. It is IMPORTANT that Craftmatic receive confirmation that you have received this notification. To assist us in this regard please check or select one of the web links provided below. Selection the first option will automatically notify us that you have received ad read this notification. Selecting the second option will take you to a different webpage where you will be asked to answer a series of questions. That webpage will provide instruction for providing the required information. Proceed by selecting the Web Link Option 1 or 2 as applicable: 1. ___I received this notice and the information above does not apply. 2. ___I received this notice and will call in reference to the information above. Please Note: If your beds are installed with both rubberized castor cups and dual beds have been secured you must still notify us that you have received this notice and understand what is requested of you. RESOLUTION: Issue 1. If you require to be supplied with rubberized castor cups or need as
Initiated
2017-03-24
Root cause
Component design/selection
Distribution
US Nationwide distribution - product was not distributed outside of the United States
Recall number
Z-1499-2022
Recall status
Open, Classified
CDRH record ID
192388
FDA record created
2022-08-04
Recall event number
89765

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗