Z-1500-2020: Rheumatoid Factor (RF) Kit for use on SPAPLUS, for in-vitro diagnostic use, Product code: LK151.S.A. , UD

Recall record Z-1500-2020 for The Binding Site Group, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1500-2020
Company in FDA record
The Binding Site Group, Ltd.
FDA record creation date
Mar 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It has been identified that the rheumatoid factor (RF) calibrator values over recover in comparison to the reference material RF serum. This could result in reporting falsely abnormal results which may lead to unnecessary additional serological testing.

Source facts

Action
It has been identified that the Rheumatoid Factor calibrator values, for LK151.S.A lot 428217, over recover in comparison to the reference material RF serum, NIBSC. To address this, the calibrator concentration values for the calibrators (SC151.A - E lot 448765) supplied in this kit have been adjusted to align more closely with the reference material. Due to the adjustment of the calibrator values, users may observe a decrease of approximately 21% in patient results in comparison to results obtained using previous calibrator values. In the absence of adopting these changes identified in the field safety notice, there is a risk, when using the affected kit, of reporting falsely abnormal results which may lead to unnecessary additional serological testing. The recalling firm issued a field safety notice along with an e-back form via email to the firm's US agent in CA on 11/6/2019. The US agent was instructed to send the field safety notice and e-back form to its customer via email. The field safety notice issued provides affected customers with reassigned calibrator values as well as a reassigned target and acceptance range for the quality controls provided with kit lot 428217 of LK151.S.A. The FSN instructs customers to input the adjusted values into the assay parameters on the SPAPLUS instrument instead of those supplied on the QC certificate.
Initiated
2019-11-06
Root cause
Nonconforming Material/Component
Terminated
2022-09-08
Distribution
Distributed in CA.
Recall number
Z-1500-2020
Recall status
Terminated
CDRH record ID
179649
FDA record created
2020-03-13
Recall event number
84909

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗