Z-1500-2025: Phoroptor VRx Digital Refraction System Model Numbers: 16241

Recall record Z-1500-2025 for Reichert, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1500-2025
Company in FDA record
Reichert, Inc.
FDA classification date
Mar 31, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The head of the phoropter head could come loose and potentially detach due to a default in the assembly.

Source facts

Action
On February 28, 2025 "URGENT MEDICAL DEVICE RECALL" letters were sent to customers. PLEASE TAKE THE FOLLOWING ACTIONS IMMEDIATELY. 1. Inspect your Phoroptor head with reference to Figure 1 for looseness or rocking and Figure 2 for an excessive gap between components. Figure 3 shows the normal condition. 2. If your unit exhibits looseness or rocking or if there is an excessive gap between components, it is possible that your unit has the defect condition. Please remove it from service immediately and contact Reichert Technical Support at 1-888-849-8955 or at https://www.reichert.com/en/support/rma-request so that the unit can be inspected and, if necessary, repaired by Reichert. The repair will be completed at no cost to you. 3. If your unit shows the normal condition as in Figure 3, please install the Support Bracket when you receive it. It is essential that you install the Support Bracket for the safety of patients and operators even if the components appear tight and securely connected. Without the bracket, there could be eventual separation of the components. The Support Bracket prevents this possibility. The internal investigation into this issue has indicated the need for these corrections and resulted in a plan that includes actions designed to prevent a recurrence of this situation. Reichert prides itself on the quality of devices we manufacture. Please share this letter with all the relevant personnel in your organization. Please contact Reichert Technical Support with any questions. They can be reached Monday through Friday, 8:00AM to 5:00PM Eastern Time at 1-888-849-8955 or at https://www.reichert.com/en/support/support-and-service-inquiry.
Initiated
2025-02-28
Root cause
Process control
Distribution
Domestic: Nationwide Distribution. International: Austria, Bolivia, Canada, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Morocco, Netherlands, Norway, Qatar, Saudi Arabia, Singapore, Slovakia, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
Recall number
Z-1500-2025
Recall status
Open, Classified
CDRH record ID
212452
FDA classified
2025-03-31
Recall event number
96266

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗