Z-1501-2020: Ysio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.

Recall record Z-1501-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1501-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Mar 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the systems controls may (depending on the values being processed) initiate a correction of the tube stand position, resulting in an unexpected movement.

Source facts

Action
On February 10, 2020, the firm notified affected customers via Urgent Medical Device Correction letters. Customers were informed of the product issue, and instructed to take the following actions: When using the system, users should follow the regular cautions and instructions listed in the system operator manual, particularly in chapter Safety, section Information about unit movements. In addition, users and all accompanying personnel are to pay special attention to remain outside of the zone of potential device movement (hazardous zone), e.g., remain outside the possible movement area of longitudinal and transversal direction of the ceiling stand. Siemens Healthineers has issued a field software update that will eliminate the issue described. The resolution will be provided free of charge. The software update is already available and is being distributed. Your local Siemens Healthineers Service organization will contact you to schedule the software update for your system. To schedule an earlier appointment, you may contact the firm's service organization at 1-800-888-7436.
Initiated
2020-02-10
Root cause
Component design/selection
Terminated
2020-08-31
Distribution
US Nationwide distribution in the states of AR, CA, DC, FL, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, UT, VA and WI.
Recall number
Z-1501-2020
Recall status
Terminated
CDRH record ID
179938
FDA classified
2020-03-03
Recall event number
85029

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗