Skip to content

Public FDA record

Z-1501-2021: InCore Lapidus Disposable Kit 28mm Right, Part Number: IC-LAP-28R - Product Usage: Intended as a three-pa

Recall record Z-1501-2021 for Nextremity Solutions. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1501-2021
Company in FDA record
Nextremity Solutions
FDA record creation date
Apr 27, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass through the drill bushing, causing inability of surgeon to complete surgical procedure as intended

Source facts

Action
Nextremity Solutions issued Urgent Medical Device Recall Letter to Distributors and Risk Managers on March 26,2021 stating reason for recall, health risk and action to take: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 - Medical Device Recall Return Response and send to Corporate Quality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit facility's documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574- 371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.
Initiated
2021-03-26
Root cause
Under Investigation by firm
Terminated
2021-11-10
Distribution
AL, AK, CA, IA,IL, IN, KY ,MD. MI, NJ, NY, OR, PA, SC, TX, VA
Recall number
Z-1501-2021
Recall status
Terminated
CDRH record ID
186740
FDA record created
2021-04-27
Recall event number
87665

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗