Skip to content

Public FDA record

Z-1501-2022: Turbett Surgical Container, TS1500

Recall record Z-1501-2022 for Turbett Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1501-2022
Company in FDA record
Turbett Surgical, Inc.
FDA classification date
Aug 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Devices were subject to unapproved rework processes.

Source facts

Action
The firm distributed recall notification to hospitals via email on 04/14/2022. The recall notification identifies the reason for recall of specific TS1500 units to be that the repair process employed could restrict the customer's ability to clean and visually check the device for cleanliness. This can be identified by the customer by evidence of bioburden or particulate buildup in the seams of the device. Inadequate or incomplete cleaning of the unit may compromise the sterility of the load. The firm asks customers to immediately discontinue use of the device and complete the provided Return Response Form to Quality@Turbettsurgical.com. Turbett will contact customers to coordinate the replacement of the unit and correction of the recalled device. Customers with any questions can call 585-755-0133x106 from Monday through Friday, from 8:00 AM to 4:30 PM.
Initiated
2022-04-14
Root cause
Process control
Terminated
2025-02-20
Distribution
US Nationwide Nationwide: CO, FL, NC and PA.
Recall number
Z-1501-2022
Recall status
Terminated
CDRH record ID
194851
FDA classified
2022-08-04
Recall event number
90578

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗