Devices with v38 system software do not display a ''LOW BATTERY: CONNECT TO AC POWER'' message when the monitor is on backup (DC) battery power and may shut down without warning.
Source facts
Action
On August 30, 2006 the firm issued a letter dated August 2006 and titled MEDICAL DEVICE CORRECTION. The letter advises of the issue, explains how to resume power if power is shut down without warning, and advsies that a service representative will contact the consignee to make arrangements to install new software. On 10/19/06 the firm mailed an identical letter dated October 2006 to customers for units that were later identified as being subject to the recall.
Initiated
2006-08-30
Root cause
Other
Terminated
2011-03-25
Distribution
Units were distributed worldwide to hospitals and medical centers.
Recall number
Z-1502-06
Recall status
Terminated
CDRH record ID
48154
FDA classified
2006-09-13
Recall event number
36190