Skip to content

Public FDA record

Z-1502-2020: AGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Us

Recall record Z-1502-2020 for Agfa N.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1502-2020
Company in FDA record
Agfa N.V.
FDA record creation date
Mar 18, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.

Source facts

Action
AFGA issued via email On February 10, 2020, an URGENT FIELD SAFETY NOTICE letter. Letter states reason for recall, health risk, and action to take: With this letter, Agfa is informing you that the DR 800 system still can be used including for Tomosynthesis for mAs values which are larger than 0,2 mAs. In case Tomosynthesis image acquisition sequence does not stop automatically after the expected number of exposures, release the exposure button as this will stop the sequence and notify your Agfa Service contact at once. Contact your local Agfa Imaging organization 1-877-777-2432 and please reference Problem Record PRB2000387 and Vigilance Record VR0000716. Complete/Return by fax or email the URGENT FIELD SAFETY NOTICE Feedback form
Initiated
2020-02-10
Root cause
Under Investigation by firm
Terminated
2023-03-30
Distribution
US Nationwide distribution in the states of FL, VA.
Recall number
Z-1502-2020
Recall status
Terminated
CDRH record ID
179937
FDA record created
2020-03-18
Recall event number
85028

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗