Open original FDA record ↗Recall record
- FDA identifier
- Z-1502-2020
- Company in FDA record
- Agfa N.V.
- FDA record creation date
- Mar 18, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.
Source facts
- Action
- AFGA issued via email On February 10, 2020, an URGENT FIELD SAFETY NOTICE letter. Letter states reason for recall, health risk, and action to take: With this letter, Agfa is informing you that the DR 800 system still can be used including for Tomosynthesis for mAs values which are larger than 0,2 mAs. In case Tomosynthesis image acquisition sequence does not stop automatically after the expected number of exposures, release the exposure button as this will stop the sequence and notify your Agfa Service contact at once. Contact your local Agfa Imaging organization 1-877-777-2432 and please reference Problem Record PRB2000387 and Vigilance Record VR0000716. Complete/Return by fax or email the URGENT FIELD SAFETY NOTICE Feedback form
- Initiated
- 2020-02-10
- Root cause
- Under Investigation by firm
- Terminated
- 2023-03-30
- Distribution
- US Nationwide distribution in the states of FL, VA.
- Recall number
- Z-1502-2020
- Recall status
- Terminated
- CDRH record ID
- 179937
- FDA record created
- 2020-03-18
- Recall event number
- 85028