Open original FDA record ↗Recall record
- FDA identifier
- Z-1503-2014
- Company in FDA record
- Philips Medical Systems, Inc.
- FDA classification date
- Apr 24, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The footboard of Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held.
Source facts
- Action
- On March 17, 2014, consignees were notified of the recall via letter. The letter explains what the issue is, actions firms can take to avoid or minimize the occurrence of the issue, and the actions planned by Philips to correct the issue. For further information or support concerning this issue, please contact your local Philips Representative. 1-800-722-9377
- Initiated
- 2014-03-17
- Root cause
- Process control
- Terminated
- 2020-09-16
- Distribution
- Nationwide: US: AL, AR, AZ, CA, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, WY / Worldwide including Canada & Mexico
- Recall number
- Z-1503-2014
- Recall status
- Terminated
- CDRH record ID
- 126536
- FDA classified
- 2014-04-24
- Recall event number
- 67859