Skip to content

Public FDA record

Z-1504-2021: Power Wheelchair

Recall record Z-1504-2021 for Merits Holdings Co. dba Merits Health Products, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1504-2021
Company in FDA record
Merits Holdings Co. dba Merits Health Products, Inc.
FDA record creation date
Apr 28, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Seat baseplate pin has broken in the field.

Source facts

Action
A Recall notification letter titled, "MEDICAL DEVICE FIELD CORRECTION" was sent to consignees on 04/07/2021 via email and by certified mail on 04/16/2021. The firm stated they made phone call notifications on 04/01/2021. The letter instructs the consignee to Contact impacted customer using a personalized customer letter (template provided). The notification can be sent by mail, email or phone. (IF by phone, please keep a log of call. Date/Time/Result); replace the old seat plate with new one provided using the SOP provided. The field correction should take approximately 15 minutes to complete.; complete and return the Service and Repair Report (sent with new part), inspect the installation and note it on the appropriate Scooter or Power Chair Inspection Record provided; use the RA shipping label (sent with new part) and return old seat plate to Merits Health Products along with the Service and Repair Report (one filled out for each unit); and when all of the units in the possession of the consignee are finished, please return the completed Inspection Report/s to Merits by mail or email.
Initiated
2021-04-01
Root cause
Process change control
Distribution
Distributed nationwide to AL, AK, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IO, KS, KY, LA, MN, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, TN, TX, UT, VT, VA, WA, WV, WI and internationally to Canada
Recall number
Z-1504-2021
Recall status
Open, Classified
CDRH record ID
186741
FDA record created
2021-04-28
Recall event number
87673

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗