Public FDA record

Z-1504-2025: Brand Name: QIAstat-Dx Product Name: QIAstat-Dx Respiratory Panel Plus Reference Number (REF): 691224 Sof

Recall record Z-1504-2025 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1504-2025
Company in FDA record
Qiagen Sciences LLC
FDA classification date
Mar 31, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Identified faulty cartridges. If such a faulty cartridge is used for sample testing, false test results could occur.

Source facts

Action
On March 12, 2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. Actions to be taken by customer/user: 1. Check if you have remaining stock. DO NOT use it. Dispose of it immediately in accordance with your national and local safety and environmental regulations. 2. Contact QIAGEN Technical Services for a free-of-charge replacement. o If you already used cartridges from this LOT, identify the results obtained with cartridges from the affected SN. For results obtained with an affected SN: Review all results of the respective targets mentioned in Detailed description of the issue, above, to identify potentially erroneous diagnosis, except in those cases where alternative confirmation was obtained. o Review any result for which tests were performed, or treatments were given on clinical and/or epidemiological suspicion of the respective targets. 3. Review this notice with your laboratory/medical director. 4. Important: Forward this information to all individuals and departments within your organization using the above-listed kits. If you are not the end user, forward this notice to the product end user. 5. Complete the Acknowledgement of Receipt Form Actions taken by QIAGEN: QIAGEN was able to identify the root cause of the issue and has implemented immediate actions to ensure that the currently produced cartridges will not cause this issue moving forward. If you have any questions or concerns, please contact your local QIAGEN Technical Services Department through any of the following: Telephone: 800 362 7737; Email: TechService-NA@qiagen.com
Initiated
2025-03-12
Root cause
Mixed-up of materials/components
Distribution
US Distribution: AL, CA, DC, FL, IN, KS, KY, MA, ME, MN, NY, and TX.
Recall number
Z-1504-2025
Recall status
Open, Classified
CDRH record ID
213097
FDA classified
2025-03-31
Recall event number
96521

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗