False Negative Staphylococcus aureus results which could prevent infected patients from receiving antimicrobial treatment; due to the failure of the Positive Control to react within 20 seconds. The failure is due to reduced reactivity of the latex component.
Source facts
Action
The recalling firm notified distributors and end users by e-mail, faxed letters, and certified courier on 08/17/06.
Initiated
2006-08-17
Root cause
Other
Terminated
2007-03-09
Distribution
Worldwide-product was sold to distributors and end users
Recall number
Z-1505-06
Recall status
Terminated
CDRH record ID
48291
FDA classified
2006-09-16
Recall event number
35794