It was observed that the corrugated hose holder above the 3D rail on the Multix Fusion system became loose. According to the Installation Instructions the corrugated hose holder shall be afixed with transversal carriage, with Loctite 243 and Torque 10nm during system installation. Siemens initiated the update instruction to verify and secure the hose holder with Loctite 243. A PCB D802 Rev. 00 b
Source facts
Action
Affected customers were initially notified with a recall letter on/about 2/12/14 and were sent Update Instructions to affected customers.
Initiated
2014-02-13
Root cause
Process change control
Terminated
2015-03-16
Distribution
Distributed in the states of AK, PA, KY, OH, WI, MI, FL, and NY.
Recall number
Z-1505-2014
Recall status
Terminated
CDRH record ID
125728
FDA classified
2014-04-24
Recall event number
67568