Open original FDA record ↗Recall record
- FDA identifier
- Z-1506-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- May 10, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There is a potential malfunction in the connection of a Sensis EP system with Carto(R) 3 dedicated cable set (part number 14411343) with the Carto 3 system from Bionsense Webster.
Source facts
- Action
- Siemens sent a recall/correction letter dated February 22, 2012, via Update Instruction AX049/11/S of this product to all affected customers. The letter informed customers of the potential issues and provides additional information on the implemented resolution. An Update Instuction AX048/11/S will be available immediately. The update instruction provides a new version of the Carto 3 addendum to the Operator Manual and the exchange of the previous stim cables with the latest cable. For questions call 610-219-4834.
- Initiated
- 2012-02-22
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-01-24
- Distribution
- Nationwide Distribution including FL, NC, and SC.
- Recall number
- Z-1506-2012
- Recall status
- Terminated
- CDRH record ID
- 108334
- FDA classified
- 2012-05-10
- Recall event number
- 61517