- FDA identifier
- Z-1506-2021
- Company in FDA record
- Datascope Corp.
- FDA classification date
- Apr 28, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
There are cybersecurity vulnerabilities in a widely used low-level TCP/IP software library that may result in a loss of communication to the Hospital Information System/Clinical Information System (HIS/CIS).
Source facts
- Action
- URGENT MEDICAL DEVICE CORRECTION notification letters dated 3/22/21 were distributed to customers. PLEASE FORWARD THIS INFORMATION TO ALL CURRENT AND POTENTIAL CARDIOSAVE HYBRID and CARDIOSAVE RESCUE IABP USERS WITHIN YOUR HOSPITAL / FACILITY. IF YOU ARE A DISTRIBUTOR WHO HAS SHIPPED ANY AFFECTED PRODUCTS TO CUSTOMERS, PLEASE FORWARD THIS DOCUMENT TO THEIR ATTENTION FOR APPROPRIATE ACTION. Corrective Action: Datascope/Getinge is currently developing a software correction to address this issue. A Datascope/Getinge service representative will contact you to schedule the installation of the updated software. This work will be done at no cost to your facility. Please complete and sign the attached MEDICAL DEVICE CORRECTION RESPONSE FORM (page 5) to acknowledge that you have received this notification. Return the completed form to Datascope/Getinge by e-mailing a scanned copy to cardiosaveripple20.act@getinge.com or by faxing the form to 1-800-506-1644.
- Initiated
- 2021-03-22
- Root cause
- Software design
- Distribution
- Worldwide distribution - US Nationwide distribution.
- Recall number
- Z-1506-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186139
- FDA classified
- 2021-04-28
- Recall event number
- 87482