Skip to content

Public FDA record

Z-1507-2024: PHILIPS Incisive Computed tomography X-ray system Incisive CT Software Upgrade from V4.5.1.41921 to V4.5.

Recall record Z-1507-2024 for Philips North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1507-2024
Company in FDA record
Philips North America
FDA classification date
Apr 9, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Patient tabletop moved out to the home position during scan initialization, may cause operator/bystander staying by the rear of the table to collide with the moved tabletop and be injured

Source facts

Action
Philips Healthcare issued Customer Information letter on 12/24/20 via USPS. Letter states reason for recall, health risk and action to take: You may continue to use your system(s) in accordance with the intended use. Refer to ADVICE ACTIONS BY CUSTOMER/USER of the customer letter for specific details regarding the issue descriptions and advice to customers. Should the customer feel uncertain regarding this action, please contact your local Philips representative. Field Change Orders (FCO72800727) involving software upgrade will be released to correct the issue at no charge to you. The planned release date is in Q1 2021. A Philips Field Service Engineer will contact you to schedule the upgrade at your site.
Initiated
2020-12-24
Root cause
Software design
Distribution
Nationwide including Puerto Rico Foreign: Albania Argentina Australia Austria Bangladesh Bolivia Brazil Bulgaria China Czech Republic Denmark Ecuador Egypt France Germany Greece Hungary India Indonesia Iraq Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kuwait Latvia Lebanon Libya Malaysia Myanmar Nepal Netherlands Norway Pakistan Palestine, State of Peru Philippines Poland Puerto Rico Qatar R¿union Romania Russian Federation Saudi Arabia Senegal Serbia Spain Sweden Switzerland Thailand Tunisia Turkey United Arab Emirates United Kingdom Uzbekistan Viet Nam
Recall number
Z-1507-2024
Recall status
Open, Classified
CDRH record ID
206606
FDA classified
2024-04-09
Recall event number
94229

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗