Z-1507-2025: BD 4 Fr Single-Lumen PowerPICC Catheters, REF: 3174155, CK000540, 3174118, CK000483, 3174335, 3174355, CK

Recall record Z-1507-2025 for Bard Access Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1507-2025
Company in FDA record
Bard Access Systems, Inc.
FDA classification date
Apr 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Catheters indicated for short-or-long term access to the central venous system have tubing that exceeded melt flow index specification (higher), which could lead to increased material fatigue leaks, which could lead to infiltration, extravasation, discomfort, phlebitis, bleeding, air embolism, foreign body embolism, infection and interruption to therapy.

Source facts

Action
On 3/11/2025, recall notices were distributed to customers and distributors who were asked to do the following: Actions if catheter damage is suspected: 1. Stop any infusion if catheter damage is suspected. 2. Follow your institution's guidelines for catheters with suspected damage. 3. If catheter fracture is confirmed, catheter should be removed and an alternative route for access should be obtained. 4. Report any catheter replacements due to leakage to firm's Complaint Center. Please Take the Following Actions: 1. Destroy all unused product following your institution's process for destruction. 2. If affected device was previously used on a patient without incident, no further action is necessary. 3. Recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. 4. Complete and return Customer Response Form via email to Email: BDRC44@BD.com Distributors asked to identify all customers within distribution network that purchased any affected product. Provide a copy of the attached customer letter to all customers to advise them of this recall Actions Taken by Firm: 2. Firm is notifying users that adhesive-backed securement systems are recommended. 3. Firm is notifying users that if they do use compression-style securement systems, they must be sized appropriately to accommodate the taper region. 4. Firm has been updating the product IFU with cautions related to the use of compression-style securement systems around the taper region of the PICC. 5. Firm is notifying customers to insert the catheter as close as possible to the zero mark per the IFU. For further assistance, contact: North American Regional Complaint Center: 1-844-823-5433 Mon-Fri 8:00am and 5:00pm CT or productcomplaints@bd.com. On 5/15/2025 additional recall notices were distributed to customers informing them of additional affected devices.
Initiated
2025-03-11
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution including in the states of WI, NC, TX, NY, CA, MN, MT, RI, MA, OH, CT, SC, IL, CO, VA, NJ, GA, UT, FL, PA, IN, MD, KY, AZ, MI, WA, HI, MO, VT, AL, NM, TN, DE, LA, IA, NH, ID, DC, WV, KS, NE, SD, ND, WY, AK, OK, OR, AR and the countries of Andorra, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, French Polynesia, French Guayana, Germany, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Luxembourg, Martinique, Mexico, Netherlands, New Caledonia, New Zealand, Norway, Pakistan, Poland, Portugal, Qatar, Republic of Korea, Reunion, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Viet Nam.
Recall number
Z-1507-2025
Recall status
Open, Classified
CDRH record ID
212851
FDA classified
2025-04-16
Recall event number
96426

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗