Skip to content

Public FDA record

Z-1508-2021: Change Healthcare Enterprise Viewer

Recall record Z-1508-2021 for CHANGE HEALTHCARE CANADA COMPANY. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1508-2021
Company in FDA record
CHANGE HEALTHCARE CANADA COMPANY
FDA record creation date
Apr 28, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Change Healthcare has identified an intermittent software defect which may result in an anchor study failing to display.

Source facts

Action
On 3/15/21 the firm sent an Urgent Field Safety Notice to its customers with the following instructions: Immediate actions to take to reduce the risk to patient When the issue occurs, the initial anchor images are populated in the viewport. The user can select the series identifier in the viewport to display other series for the study. If the initial anchor images in the viewport have been replaced with reference images, the user does not have the ability to view the anchor series. However, the user can reopen the study from the EHR or perform a specific search in CHEV to allow the anchor study series to display in the viewport. In addition, the user can change the timeline filter to omit the scheduled study, which would allow the anchor study to display in the timeline. Product updates that will address the issue Change Healthcare has developed a software update to correct this issue. Recommended actions to take on potentially affected data We strongly encourage customers receiving this Urgent Field Safety Notice to immediately contact Change Healthcare Customer Support to: 1) Acknowledge they have read and understood this Urgent Field Safety Notice; and, 2) Arrange the installation of software update.
Initiated
2021-03-17
Root cause
Software design
Distribution
Domestic Distribution only. IN, KY, CA, NY, WA, IL, IA, FL, TX, WA, MI, HI, AK, WI, WY, NE, AZ, OH, PA
Recall number
Z-1508-2021
Recall status
Open, Classified
CDRH record ID
186406
FDA record created
2021-04-28
Recall event number
87616

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗