Z-1508-2022: The Cholestech LDX System combines enzymatic methodology and solidphase technology to measure total chole

Recall record Z-1508-2022 for Alere San Diego, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1508-2022
Company in FDA record
Alere San Diego, Inc.
FDA record creation date
Aug 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).

Source facts

Action
On 6/24/22, correction notices were mailed and emailed to customers who were asked to take the following actions: 1) Test results from patients with bilirubin levels above 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated) should be verified using another test method. 2) Complete and return the Customer Reply Form. 3) If you have forwarded the products listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. 4) Retain this letter for your laboratory records. Customers with questions are encouraged to contact Technical Service at 877-308-8289.
Initiated
2022-06-24
Root cause
Device Design
Distribution
US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR, HK, BR, CO, AR, CA
Recall number
Z-1508-2022
Recall status
Open, Classified
CDRH record ID
194622
FDA record created
2022-08-06
Recall event number
90508

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗