Skip to content

Public FDA record

Z-1508-2024: EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecula

Recall record Z-1508-2024 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1508-2024
Company in FDA record
Qiagen Sciences LLC
FDA classification date
Apr 10, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost

Source facts

Action
Qiagen issued Urgent Medical Device Correction letter to Lab Director/Manager, Medical Director, Risk Manager, Safety Officer on 3/4/24 via email. Letter states reason for recall, health risk and action to take: Complete the Acknowledgment of Receipt promptly and return to quality.communications@qiagen.com. Should the error described above occurs, please perform the following actions: o Switch off the instrument and disconnect from the power supply. o Remove any onboard samples and perform decontamination procedure of the instruments as described in the user manual. o Call your local QIAGEN Technical Service or the Commercial Partner responsible for QIAGEN products in your country. If you have any questions or concerns, please contact your local QIAGEN Technical Service Department through any of the following: " Telephone: (800) 362-7737 " Email: TechService-NA@qiagen.com
Initiated
2024-03-04
Root cause
Under Investigation by firm
Distribution
Nationwide Foreign: AT, AU, BE, CH, CR, CZ, DE, DK, ES, FR, GB, HU, IT, JO, NL, NO, QA, SA, SE, SI, TH
Recall number
Z-1508-2024
Recall status
Open, Classified
CDRH record ID
206743
FDA classified
2024-04-10
Recall event number
94254

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗