Skip to content

Public FDA record

Z-1511-2025: Cordis, REF: C09060ML, Smart Control Vascular Stent System ILIAC, 9mm X 60mm 6F, SterileEO, Rx Only

Recall record Z-1511-2025 for Cordis US Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1511-2025
Company in FDA record
Cordis US Corp
FDA classification date
Apr 3, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

product mix-up; Vascular stent labeled as one size but contains a different size.

Source facts

Action
On March 3, 2025, Cordis issued an "URGENT MEDICAL DEVICE RECALL" notification via FedEx. Cordis asked consignees to take the following actions: 1. Read this Urgent Medical Device Recall letter. 2. Immediately check your inventory for any units from the affected lot in your possession. Identify and quarantine any units from the identified lot in a manner that ensures the affected product will not be used. Check all storage and usage locations. 3. Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. 4. Review, complete, sign and return the enclosed Acknowledgement Form directly to Cordis at the fax number on the form or email to: GMB-CordisFieldAction@cordis.com 5. Return any affected product to the address listed on the form, with reference to your Customer Number which is listed on the form. 6. Maintain awareness of this notice until all affected product has been returned to Cordis and keep a copy of this notice with the affected product. For questions related to the recall and/or acknowledgement form that are not adequately addressed in this letter, please contact Cordis QA at: GMB-CordisFieldAction@cordis.com or call Cordis Customer Support 1-800-327-7714 Option 2, Monday through Friday from 8:00 AM to 8:00 PM EDT.
Initiated
2025-03-04
Root cause
Employee error
Distribution
US Distribution to states of: AR, AZ, FL, NJ, TN; and OUS Foreign country of: Japan
Recall number
Z-1511-2025
Recall status
Open, Classified
CDRH record ID
212827
FDA classified
2025-04-03
Recall event number
96418

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗