Skip to content

Public FDA record

Z-1512-06: Boston Scientific Cutting Balloon Ultra2 Monorail Device, Boston Scientific, One Boston Scientific Plac

Recall record Z-1512-06 for Boston Scientific. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1512-06
Company in FDA record
Boston Scientific
FDA classification date
Sep 23, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Lack of assurance of sterility (pre-sterilization bioburden limits exceeded)

Source facts

Action
On July 21, 2006, Boston Scientific initiated a voluntary recall of 45 lots/batches of Cutting Balloon Ultra Monorail Devices due to their determination that the lots do not meet their sterilization requirements. There is a remote possibility of infection in the patient. An infection, if it occurred, could be recognized and treated effectively in these patients. Boston Scientific is not aware of any complaints related to this product issue. Letters dated July 21, 2006, were sent to their customers with instructions to cease further distribution, complete and return a tracking form, and product return instructions.
Initiated
2006-07-21
Root cause
Other
Terminated
2007-02-03
Distribution
Class II Recall - Nationwide distribution ---- including states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV, and WY.
Recall number
Z-1512-06
Recall status
Terminated
CDRH record ID
47236
FDA classified
2006-09-23
Recall event number
35939

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗