Z-1512-2024: Drger Perseus A500 Anesthesia Workstation

Recall record Z-1512-2024 for Draeger Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1512-2024
Company in FDA record
Draeger Medical, Inc.
FDA classification date
Apr 18, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Unexpected shutdown while operating on battery power.

Source facts

Action
Consignees were sent an Urgent Medical Device Recall notification by mail on 3/20/24. The notice informs consignees of the possibility of device shutdown with possible loss of ventilation, which consignees are instructed to either use an emergency ventilation bag or apply manual ventilation with the Perseus with the provided instructions. Consignees are to perform a battery test on all affected devices in their inventory per the provided instructions in the recall notification. If the tested unit(s) fails the battery test, consignees are to contact Drager Service Technical Support between 8:00 AM and 8:00 PM at 1-800-437-2437 (option 2, option 2, option 2) to arrange replacement of the device's battery. Consignees are also instructed to avoid short discharging/charging cycles of devices; if this cannot be accommodated device batteries are to be tested every three months. A supplemental IFU has been released to include the additional maintenance instructions outlined in the recall notification and can be requested from the firm. Consignees with any questions can contact Michael Kelhart between 8:00 AM to 4:30 PM EST at 267-664-1131 or mike.kelhart@draeger.com.
Initiated
2024-03-20
Root cause
Labeling design
Distribution
Domestic: Nationwide Distribution; Foreign: Angola, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-se- S, Bulgaria, Canada, Chili, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, French Guiana, Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Liechtenstein, Lithuania, Malaysia, Maldives, Mauritius, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, Uganda, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.
Recall number
Z-1512-2024
Recall status
Open, Classified
CDRH record ID
206777
FDA classified
2024-04-18
Recall event number
94269

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗