Open original FDA record ↗Recall record
- FDA identifier
- Z-1512-2025
- Company in FDA record
- Qapel Medical Inc.
- FDA classification date
- Apr 7, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Aspiration catheter distal tip features and characteristics may not be in the scope of FDA clearance; tip detachment, vessel rupture, and vasospasm could occur.
Source facts
- Action
- On 2/26/2025, removal and discontinuation notices were mailed to customers were asked to do the following: 1) Quarantine affected devices and return to the firm. 2) If you have provided this product to others, let the firm know so they can notify them of this action and retrieve any remaining product. 3) Complete and return the Acknowledgement and Response Card via email to orders@qapelmedical.com If you have any questions regarding this letter, please contact customer service for further information at orders@qapelmedical.com, or by phone at 510-738-6255 between the hours of 8am - 5pm PST.
- Initiated
- 2025-02-26
- Root cause
- Device Design
- Distribution
- US: CA, IL, NE, TN, TX, NJ, NC, SD, FL, MI, PA, GA, AL, IN, NY, WI, AZ, MN, LA, OH, OK, SC, NV, ND, ID, MO, CT, WA, MA. OUS: QATAR, United Arab Emirates, Republic of Kazakhstan
- Recall number
- Z-1512-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213027
- FDA classified
- 2025-04-07
- Recall event number
- 96492