The head section wheel assembly may flex outward from the cot because the head section wheel bolts, which hold the assembly upright, may back out.
Source facts
Action
Consignees were notified of the recall via letter dated 8/29/06. Included with the letter were replacement bolts for each cot and installation instructions, but the customer was given the option of contacting the recalling firm for installation.
Initiated
2006-08-29
Root cause
Other
Terminated
2007-04-18
Distribution
Worldwide distribution ---- including USA and countries of Canada, China, Germany and Switzerland.
Recall number
Z-1513-06
Recall status
Terminated
CDRH record ID
48076
FDA classified
2006-09-22
Recall event number
36149