Open original FDA record ↗Recall record
- FDA identifier
- Z-1513-2014
- Company in FDA record
- Brainlab AG
- FDA classification date
- Apr 25, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When using multiple radiation treatment targets within a single plan, the ExacTrac v.6.0.3 might position the patient to an unintended radiation treatment point. As a result, there is a possibility that the radiation treatment dose may be delivered to the unintended target position.
Source facts
- Action
- A FIELD SAFETY NOTICE / PRODUCT INFORMATION letters, dated February 24, 2014, were sent to all direct accounts. The letter identified the affected product, the reason for the recall, and included instructions for direct accounts to take while a permanent fix can be deployed. The letter also advised that a software update will be made available starting the end of April 2014. Customers with questions or concerns can contact Brainlab at 800-597-5911 or via e-mail at us.support@brainlab.com.
- Initiated
- 2014-02-24
- Root cause
- Software design
- Terminated
- 2017-04-24
- Distribution
- Worldwide Distribution -- USA, including the states of CA, IN, LA, MA, OH, TX, and WI, and the countries of Australia, China, France, Germany, Japan, Netherlands, and the United Kingdom.
- Recall number
- Z-1513-2014
- Recall status
- Terminated
- CDRH record ID
- 126143
- FDA classified
- 2014-04-25
- Recall event number
- 67719