Open original FDA record ↗Recall record
- FDA identifier
- Z-1513-2022
- Company in FDA record
- Spacelabs Healthcare, Inc.
- FDA record creation date
- Aug 11, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.
Source facts
- Action
- On 05/31/22, correction notices were mailed to customers who were asked to take the following actions that can be taken to minimize risk to patients or users: Weigh the benefits versus the risks when deciding whether to continue to use the alarm escalation feature within XTR 1.4. Suggestions from Spacelabs to mitigate the risk: - Set the priority to the same level as the escalation max on the product. This will ensure that the highest priority alarm is triggered for the technical alarm. - Do not run the XTR 1.4 for more than 24 days consecutively. Restarting the XTR after 24 days will mitigate the issue for another 24 days. - The recalling firm is working on a 1.4.1 release of XTR which will address this defect. For additional information or technical assistance, please contact: Technical Support, 1-800-522-7025 and select 2
- Initiated
- 2022-05-31
- Root cause
- Software Design Change
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of NY, AR, CO, FL, NJ, AL, KY, TN, SC, WA, OK, MN, PA, GA, VA, CA, OH, IA, KS, MI, IN, TX, MT, MO, ME, LA, IL, SD, MA, OR, ID, MD, AZ, AK and the countries of MB, ON, FR, QC, BC, IS, CT, RM.
- Recall number
- Z-1513-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194648
- FDA record created
- 2022-08-11
- Recall event number
- 90529