Skip to content

Public FDA record

Z-1513-2022: Xhibit Telemetry Receiver, Model: 96280

Recall record Z-1513-2022 for Spacelabs Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1513-2022
Company in FDA record
Spacelabs Healthcare, Inc.
FDA record creation date
Aug 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.

Source facts

Action
On 05/31/22, correction notices were mailed to customers who were asked to take the following actions that can be taken to minimize risk to patients or users: Weigh the benefits versus the risks when deciding whether to continue to use the alarm escalation feature within XTR 1.4. Suggestions from Spacelabs to mitigate the risk: - Set the priority to the same level as the escalation max on the product. This will ensure that the highest priority alarm is triggered for the technical alarm. - Do not run the XTR 1.4 for more than 24 days consecutively. Restarting the XTR after 24 days will mitigate the issue for another 24 days. - The recalling firm is working on a 1.4.1 release of XTR which will address this defect. For additional information or technical assistance, please contact: Technical Support, 1-800-522-7025 and select 2
Initiated
2022-05-31
Root cause
Software Design Change
Distribution
Worldwide distribution - US Nationwide distribution in the states of NY, AR, CO, FL, NJ, AL, KY, TN, SC, WA, OK, MN, PA, GA, VA, CA, OH, IA, KS, MI, IN, TX, MT, MO, ME, LA, IL, SD, MA, OR, ID, MD, AZ, AK and the countries of MB, ON, FR, QC, BC, IS, CT, RM.
Recall number
Z-1513-2022
Recall status
Open, Classified
CDRH record ID
194648
FDA record created
2022-08-11
Recall event number
90529

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗