Skip to content

Public FDA record

Z-1514-2020: IMMULITE 2000 Systems E2 Estradiol Catalog #: US: L2KE22(D) (200 tests) and L2KE26(D) (600 tests) and OUS

Recall record Z-1514-2020 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1514-2020
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Feb 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to high discordant Estradiol results which.potentially affects all patient populations, A falsely elevated estradiol level could lead a clinician to misinterpret a patient as pre-menopausal when truly post- menopausal. This may lead to delayed initiation of a potentially beneficial drug and/or administration of an unnecessary drug in the treatment for hormone receptor positive advanced or metastatic breast cancer.

Source facts

Action
An Urgent Medical Device Correction (UMDC) IMC20-01.A.US was sent to US customers and an Urgent Field Safety Notice (UFSN) IMC20-01.A.OUS was sent to OUS customers in February 2020. The UMDC and UFSN advise customers of the potential for high discordant Estradiol results in some samples on the IMMULITE systems. Siemens Healthineers has determined kit lots 501 and above released in July 2018 for the IMMULITE systems are potentially affected. The UMDC instructs US customers to discontinue use of and discard Estradiol kits currently in their inventory. Customers will also be asked to do a look back if the impacted Estradiol kit lots were used to assess the menopausal status of a female while determining therapy for hormone receptor positive advanced or metastatic breast cancer. The UFSN instructs OUS customers to review the letter with their Medical Director to consider if a retrospective review of patient samples is necessary. Customer will also be instructed to not use any current inventory in their laboratory on patient samples used to assess menopausal status while determining therapy for hormone receptor positive advanced or metastatic breast cancer. These samples will need to be tested using an alternate methodology. Customers may continue to use the kits currently in their inventory and report results on other patient populations. Additionally, if a discordant high result for Estradiol is suspected, customers are advised to follow their established internal procedures to investigate the issue.
Initiated
2020-02-05
Root cause
Under Investigation by firm
Terminated
2021-11-18
Distribution
Worldwide distribution - US Nationwide in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, KY, LA, MA, MD, MI, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI. SC, TN, TX, UT, VA, WA, WI, WY, and countries of AE, AF, AL, AO, AR, AT, AU, AZ, BD, BE, BO, BG, BH, BR, CA, CH, CO, CL, CN, CO, CR, CY, DE, DO, DZ, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, IE, IN, IR, IS, IT, IQ, JO, KE, KR, LB, LT, LV, LY, ME, MK, MX, NL, NO, PA, PE, PH, PK, PL, PS, PT, PY, QA, RO, RU, SA, SE, SK, SL, SV, SY, TH, TJ, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XK, XS, ZA
Recall number
Z-1514-2020
Recall status
Terminated
CDRH record ID
179698
FDA classified
2020-02-24
Recall event number
84947

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗