Open original FDA record ↗Recall record
- FDA identifier
- Z-1514-2022
- Company in FDA record
- Olympus Corporation of the Americas
- FDA record creation date
- Aug 11, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment
Source facts
- Action
- Olympus issued Urgent Medical Device Recall Letter dated July 7, 2022. Letter states reason for recall, health risk and action to take: 1. Immediately locate the generator with the serial numbers listed in this communication. The serial number is located on the back of the unit. 2. Your sales representative will work with you to arrange for the return of your unit and assist with initiating an order for a replacement unit. 3. Olympus requests that you acknowledge receipt of this letter by accessing the Olympus recall portal. Go to https://olympusamerica.com/recall. Enter the recall number 0410 and provide your contact information as indicated in the portal. 4. If you have distributed these devices outside your facility, please notify your customers of this removal action immediately. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Contact (647) 999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
- Initiated
- 2022-07-07
- Root cause
- Process change control
- Distribution
- US Nationwide distribution in the states of AL, NM.
- Recall number
- Z-1514-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194859
- FDA record created
- 2022-08-11
- Recall event number
- 90585