Skip to content

Public FDA record

Z-1516-2022: Aeos Robotic Digital Microscope with software versions 2.7 or earlier Article Number: PV010

Recall record Z-1516-2022 for Aesculap Implant Systems LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1516-2022
Company in FDA record
Aesculap Implant Systems LLC
FDA record creation date
Aug 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Application error problems: Rebalance scope message: 1.Rebalancing process 2.Usage of Robot Recovery: 3. Autofocus not available:

Source facts

Action
The firm, Aesculap Inc., issued "URGENT MEDICAL DEVICE CORRECTION" letter dated July 14, 2022. Letter states reason for recall, health risk and action to take: Address the application error concerns, the product instructions for use (IFU) have been updated and a new software version for the Aeos Robotic Digital Microscope (version 2.8.1.115) has been released. All customers are being provided the updated IFU and software. Refer to Actions Required by Aesculap Inc. Customer/User below. 1. Review the Device Correction Notification in its entirety and notify users and other relevant personnel within your organization. 2. Download the updated product IFU and ensure all users are aware of the updates. To obtain the updated IFU, please perform the following steps: a. Using your organizations approved web browser (e.g. Google Chrome, Microsoft Edge, etc.), enter the URL: https://www.aesculapusaifus.com. b. In the search field enter the Item Number: PV010 and press Search c. Click on the file AEOS Digital Microscope IFU (TA015615). Note: You will need a PDF viewer such as Adobe Acrobat to view the file. d. Print or distribute the IFU to relevant personnel. Maintain a copy in close proximity to the device for easy reference by the users. 3. Complete the Product Correction Acknowledgement form. If you have no inventory remaining, enter zero (0) on the form. 4. Return the completed Product Correction Acknowledgement form to our Quality Assurance department by e-mailing to qa-recalls@aesculapusa.com, even if the total inventory in your possession is zero (0). 5. Once we have received your acknowledgement form an Aesculap representative will contact you to arrange for qualified Aesculap personnel to update the software on your device. Report any concerns or issues to our Postmarket Surveillance team by contacting ProductQualityExcellence@bbraunusa.com or by calling 1-833-425-1464.
Initiated
2022-07-14
Root cause
Software design
Terminated
2025-02-07
Distribution
US Distribution to states of: AR, IN, NE and NC
Recall number
Z-1516-2022
Recall status
Terminated
CDRH record ID
195295
FDA record created
2022-08-11
Recall event number
90641

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗