Z-1517-2008: GE Healthcare Innova 4100 Digital Fluoroscopic Imaging System with Bolus Chasing option.

Recall record Z-1517-2008 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1517-2008
Company in FDA record
GE Healthcare
FDA classification date
Sep 29, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Incorrect dose data: GE Healthcare has found that after 6 days and 4 hours without performing a system reset or a system reboot (turning it off/on), the displayed dose data on these systems (Innova 3100 / Innova 4100 with Bolus option) may be underestimated by up to 50%. In order to avoid this issue, it is necessary to perform a system reset or reboot every day. GE Healthcare is working on a c

Source facts

Action
GE sent all customers with affected units a letter on 11/15/07 titled Urgent Medical Device Correction. In the US this was sent by a traceable means (either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX overnight Delivery). The letter was sent to 3 titles within the affected accounts; Hospital Administrator, Manager of Radiology/Cardiology and Radiologist/ Cardiologist. A correction letter identified a short term recommendation to perform a system reset or reboot every day and stated that a software correction will be installed at their site at no cost.
Initiated
2007-10-21
Root cause
Nonconforming Material/Component
Terminated
2011-12-17
Distribution
Worldwide Distribution: USA, Argentina, Australia, Canada, China, Germany, Japan, Mexico and Russia.
Recall number
Z-1517-2008
Recall status
Terminated
CDRH record ID
73740
FDA classified
2008-09-29
Recall event number
45538

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗