Z-1517-2021: Philips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices - intended for use

Recall record Z-1517-2021 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1517-2021
Company in FDA record
Philips North America Llc
FDA record creation date
Apr 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is connected o the Flex Cardio device; display of certain data from the FC2010 becomes frozen, i.e., waveforms cease sweeping and updating and the ECG, IBP, and respiration numeric values cease to update;e ECG, IBP, and respiration waveforms become flat lines and no audible alarms are produced for HR and IBP, upon start up, an unexpected non physiological ECG waveform, erratic heart rate numeric value, and non-physiological display of any other active waveforms may appear on the Boom Monitor

Source facts

Action
Philps issued Urgent Field Safety Field Notice-Medical Device Correction to or U.S. customers, the letters were sent via certified mail on 3/29/21. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: Flex Cardio System. Because the issues listed above can be promptly identified by a qualified health care professional who has reviewed this notice and is closely observing the monitored patient, the Xper Flex Cardio may continue to be used if this step is taken. Resetting the device as described in the IFU (Section 4, page 87), is likely to return the device to normal operation, which takes approximately 3-5 minutes. To reset your Flex Cardio device, close any patient cases and then turn the power switch off (see back of device for the power switch). Wait 510 seconds then turn the power switch back on. Allow the device to fully restart and restore the monitoring display. Please complete and return the Customer Reply Form included on the last page of this communication indicating your receipt and understanding of this information. Philips will provide a software update for the Xper Flex Cardio to correct all but, issue #7 above at no charge. A Philips representative will contact you when the software is available for installation. Philips has also added directions to the IFU for the Xper Flex Cardio on how to reset the device in the event that the user observes issue #7 above. An IFU addendum with the directions will be provided to all affected users when it becomes available. If you need any further information or support concerning this issue, please contact your local Philips representative: 800-669-1328 option 2, then option 3.
Initiated
2021-03-29
Root cause
Software design
Distribution
Nationwide Foreign: Argentina Australia Austria Bangladesh Belgium Bulgaria Canada Chile China Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Finland France Germany Greece Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jordan Kazakhstan Kenya Korea, Republic of Kuwait Latvia Lebanon Libya Lithuania Malaysia Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nigeria Norway Oman Pakistan Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico R¿union Romania Russian Federation Saudi Arabia Serbia Singapore Slovenia South Africa Spain Sri Lanka Sweden Switzerland Syrian Arab Republic Taiwan Thailand Tunisia Turkey United Arab Emirates United Kingdom Uzbekistan Viet Nam Yemen
Recall number
Z-1517-2021
Recall status
Open, Classified
CDRH record ID
186883
FDA record created
2021-04-29
Recall event number
87732

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗