Skip to content

Public FDA record

Z-1518-2024: Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098

Recall record Z-1518-2024 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1518-2024
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Apr 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.

Source facts

Action
On 3/13/24, Field Safety Notices were mailed and emailed to customers who were asked to do the following: 1) To prevent a pinch injury, please use two people when attempting to move the EGM in an unpowered state. 2) Inform necessary hospital personnel about this issue and affected personnel when the requested actions have been completed. 3) Complete and return the response form via email to Recalls@intusurg.com A firm representative will schedule a site visit to perform the inspection of the Link 2 EGM belts of the affected population of SP systems. If you need further information or support concerning this Medical Device Notification, please contact your Clinical Sales Representative or contact Firm Customer Service: - North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com. - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 AM to 6 PM CET) or eucs@intusurg.com - South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) or support.korea@intusurg.com - Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST) or csjapan@intusurg.com
Initiated
2024-03-13
Root cause
Nonconforming Material/Component
Distribution
US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV, MS, TX, KS. OUS: China, France, Japan, South Korea
Recall number
Z-1518-2024
Recall status
Open, Classified
CDRH record ID
206739
FDA classified
2024-04-11
Recall event number
94252

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗