Skip to content

Public FDA record

Z-1519-2018: Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 t

Recall record Z-1519-2018 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1519-2018
Company in FDA record
Roche Diagnostics Corporation
FDA record creation date
Apr 25, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The previous version of method sheets for the Acetaminophen assay on the COBAS INTEGRA 400 plus/800/800 CTS analyzers and the MODULAR ANALYTICS P module are missing the following statement in the Limitations-interference section: "In very rare cases, gammopathy, in particular type IgM (Waldenstrm s macroglobulinemia), may cause unreliable results."

Source facts

Action
On January 2, 2018 an URGENT MEDICAL DEVICE CORRECTION letter was issued to customers with notification to the issue and information regarding the updated method sheets and that they were available for use. The COBAS INTEGRA 400 plus/800/800 CTS was updated from affected method sheet version 3.0 to 4.0. The MODULAR ANALYTICS P was updated from affected method sheet version 4.0 to 5.0 No product return is required. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Initiated
2018-01-02
Root cause
Error in labeling
Terminated
2019-04-25
Distribution
Nationwide Distribution
Recall number
Z-1519-2018
Recall status
Terminated
CDRH record ID
162582
FDA record created
2018-04-25
Recall event number
79526

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗