Open original FDA record ↗Recall record
- FDA identifier
- Z-1520-2008
- Company in FDA record
- Datex - Ohmeda, Inc
- FDA classification date
- Sep 30, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
User was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to incorrect overall dose management, resulting in unnecessary exposure and can be a contributing factor for skin injury. The alarm can be heard in the control room.
Source facts
- Action
- GE Field Engineers are visiting all affected customers to apply updated software. For assistance contact Datax-Ohmeda, Inc., at 1-608-221-1551.
- Initiated
- 2006-08-17
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2009-06-09
- Distribution
- within US: AL, AR, AZ, CA,CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico. OUS to include: Algeria, Australia, Belgium, Brazil, Canada, China, Columbia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Ireland, Italy, Japan, Korea, Kuwait, Latvia, Malaysia, Mexico, New Zealand, Norway, Russia, Saudi Arabia, Singapore, Spain, Switzerland, Syria, Taiwan, Turkey, and United Kingdom
- Recall number
- Z-1520-2008
- Recall status
- Terminated
- CDRH record ID
- 72442
- FDA classified
- 2008-09-30
- Recall event number
- 48977