Skip to content

Public FDA record

Z-1520-2021: Grace Medical REF S-3500 *** COX BONE DUST COLLECTOR *** 2 stainless steel filters - Product Usage: inte

Recall record Z-1520-2021 for Oto Med Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1520-2021
Company in FDA record
Oto Med Inc
FDA record creation date
Apr 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During internal Age Testing, holes were found in the sterile Tyvek pouches of the devices.

Source facts

Action
On 12/07/2020, the firm sent an "URGENT: PRODUCT RECALL NOTICE" Letter via FedEx Overnight (U.S.) and email (O.U.S.) to customer informing them that there is a potential for a hole to develop in the packaging during handling/shipment. Customers are instructed to: 1) Check their inventory for any holes or damage to the pouches, and disposition as specified below: -My inventory has been checked and included none of the affected products. -My inventory has been checked and has no damaged pouches; The product is not being returned. -My inventory has been checked and the product listed is being returned. If returning products, please obtain a Returned Authorization (RA) number. The RA number may be obtained from Customer Service (866-472-2363). In addition, the Response Form must be completed and returned to Grace Medical Group at RECALL COORDINATOR, OTO-MED, INC. c/o GRACE MEDICAL GROUP, INC., 8500 Wolf Lake Drive, Suite 110, Memphis, TN 38133, or by fax to 901-386-0950 or email at info@gracemedical.com.
Initiated
2020-12-07
Root cause
Package design/selection
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of MN, IL, PA, WI, GA, SD, FL, AR, MA, AZ, TX, MO, NC, CA, UT, DE, OR, IN, CO, AL, OH, NY, WA, DC, TN, NH, VT, VA, SC, MI, WI, MS, IA, NJ, MD, KS, CT, LA, ME, KY, HI, TN, RI, NV, ID and the countries of Argentina, Brazil, Colombia, Chile, Hong Kong, South Africa, United Arab Emirates, Taiwan, Canada, India.
Recall number
Z-1520-2021
Recall status
Open, Classified
CDRH record ID
186334
FDA record created
2021-04-30
Recall event number
87577

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.