Open original FDA record ↗Recall record
- FDA identifier
- Z-1520-2021
- Company in FDA record
- Oto Med Inc
- FDA record creation date
- Apr 30, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During internal Age Testing, holes were found in the sterile Tyvek pouches of the devices.
Source facts
- Action
- On 12/07/2020, the firm sent an "URGENT: PRODUCT RECALL NOTICE" Letter via FedEx Overnight (U.S.) and email (O.U.S.) to customer informing them that there is a potential for a hole to develop in the packaging during handling/shipment. Customers are instructed to: 1) Check their inventory for any holes or damage to the pouches, and disposition as specified below: -My inventory has been checked and included none of the affected products. -My inventory has been checked and has no damaged pouches; The product is not being returned. -My inventory has been checked and the product listed is being returned. If returning products, please obtain a Returned Authorization (RA) number. The RA number may be obtained from Customer Service (866-472-2363). In addition, the Response Form must be completed and returned to Grace Medical Group at RECALL COORDINATOR, OTO-MED, INC. c/o GRACE MEDICAL GROUP, INC., 8500 Wolf Lake Drive, Suite 110, Memphis, TN 38133, or by fax to 901-386-0950 or email at info@gracemedical.com.
- Initiated
- 2020-12-07
- Root cause
- Package design/selection
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of MN, IL, PA, WI, GA, SD, FL, AR, MA, AZ, TX, MO, NC, CA, UT, DE, OR, IN, CO, AL, OH, NY, WA, DC, TN, NH, VT, VA, SC, MI, WI, MS, IA, NJ, MD, KS, CT, LA, ME, KY, HI, TN, RI, NV, ID and the countries of Argentina, Brazil, Colombia, Chile, Hong Kong, South Africa, United Arab Emirates, Taiwan, Canada, India.
- Recall number
- Z-1520-2021
- Recall status
- Open, Classified
- CDRH record ID
- 186334
- FDA record created
- 2021-04-30
- Recall event number
- 87577