Z-1520-2025: Tandem Mobi Insulin Pump with Interoperable Technology

Recall record Z-1520-2025 for Tandem Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1520-2025
Company in FDA record
Tandem Diabetes Care, Inc.
FDA classification date
Apr 11, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.

Source facts

Action
On February 28, 2025 Tandem Diabetes Care contacted each affected customer by phone concerning a Urgent Medical Device Correction. On March 11, 2025, Tandem followed up with a "Urgent Medical Device" Recall notification via E-Mail to affected consignees. Tandem ask consignees to take the following action: 1. Turn off Control-IQ+ immediately. You may continue to use your pump without Control-IQ active. Your pump will deliver insulin based on your programmed active personal profile and will not automatically adjust insulin delivery. 2. Please refer to the user guide provided with your pump for more detailed instructions. 3. Continue using your Tandem pump as described in the User Guide. 4. Pay attention to all system alerts and alarms. 5. Always carry back-up supplies.
Initiated
2025-02-27
Root cause
Software Design Change
Distribution
US: AZ, CA, CO, CT, FL, HI, IA, ID, IL, IN, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, PA, TX, UT, VA, WA, WI OUS: N/A
Recall number
Z-1520-2025
Recall status
Open, Classified
CDRH record ID
212817
FDA classified
2025-04-11
Recall event number
96417

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗