Open original FDA record ↗Recall record
- FDA identifier
- Z-1521-06
- Company in FDA record
- Kerr Corp
- FDA classification date
- Sep 23, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Mis-pack/mis-label: The OptiBond Self Etch Primer and OptiBond Solo Plus Adhesive in the chambers were mis-packaged/mis-labeled in reverse order. Application of these products in reverse order may result in compromised bond strength.
Source facts
- Action
- A total of thirty-four (34) consignees were sent the recall communication via USPS 1st class mail on August 29, 2006. There will be an additional five (5) letters that will be sent to consignees in Europe once the recall communication is translated into their respective languages. The consignees were instructed to complete the Return Form and return any affected product in their inventory. They were also requested to recover any affected product that they may have shipped to their customers. Periodic effectiveness checks will be conducted in order to follow up with non-responders.
- Initiated
- 2006-08-29
- Root cause
- Other
- Terminated
- 2012-08-03
- Distribution
- World wide; USA and countries of Australia, Belgium, Canada, China, France, Sweden and United Kingdom
- Recall number
- Z-1521-06
- Recall status
- Terminated
- CDRH record ID
- 48183
- FDA classified
- 2006-09-23
- Recall event number
- 36200