Skip to content

Public FDA record

Z-1521-2008: GE Healthcare Innova 3100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System)

Recall record Z-1521-2008 for Datex - Ohmeda, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1521-2008
Company in FDA record
Datex - Ohmeda, Inc
FDA classification date
Sep 30, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room).

Source facts

Action
An Urgent Medical Device Correction letter, dated June 2008 was sent to customers The letter described the issue, affected products and provided short term user recommendations. GE developed a new software release that corrects this condition and will launch the corrective action to upgrade all affected systems.
Initiated
2008-05-19
Root cause
Software design
Terminated
2016-06-30
Distribution
Worldwide Distribution - AL, AZ, CA, FL,, GA, HI, IA,IL, IN, KY, LA,MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV, Puerto Rico. OUS to include: Australia, Byelarus, China, Denmark, France, Germany, Greece, Honduras, Iran, Iraq, Italy, Japan, Latvia, Lebanon, Mexico, Poland, Russia, Saudi Arabia, Turkey, United Kingdom, and Venezuela.
Recall number
Z-1521-2008
Recall status
Terminated
CDRH record ID
73763
FDA classified
2008-09-30
Recall event number
49538

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗