Open original FDA record ↗Recall record
- FDA identifier
- Z-1521-2011
- Company in FDA record
- Vascular Solutions, Inc.
- FDA classification date
- Mar 3, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Investigations of a recent Device Experience Report has made us aware of a potential problem with our Vari-Lase WireFiber, Model 7137, manufactured with the following lot numbers: 5500898 and 550896. Recently it was reported that the endovenous laser fiber cage and guidewire tip of the WireFiber became separated. It is possible that separation could occur during patient procedure and has the p
Source facts
- Action
- Vascular Solutions sent an "Urgent Medical Device Recall" letter dated January 10, 2011. The letter described the issue and the product affected. Advised consignees to immediately check their inventory , place affected product in a secure location, and to complete and return the Customer Inventory Form. For questions regarding this recall call 763-656-4210.
- Initiated
- 2011-01-10
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-05-03
- Distribution
- Nationwide Distribution - USA including CT, MA, MN, NY, VA
- Recall number
- Z-1521-2011
- Recall status
- Terminated
- CDRH record ID
- 97562
- FDA classified
- 2011-03-03
- Recall event number
- 57847