Open original FDA record ↗Recall record
- FDA identifier
- Z-1521-2013
- Company in FDA record
- Applied Medical Resources Corp
- FDA classification date
- Jun 7, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a potential for the trigger to fracture and the jaws to remain in the closed position. The likelihood of this situation to occur and result in permanent patient injury is highly unlikely;
Source facts
- Action
- Applied Medical notified customers of the recall beginning on May 13, 2013. The "URGENT: MEDICAL DEVICE RECALL" letter identified the product, problem, and actions to be taken by the customers. For questions regarding this notice, contact the firm at (949) 713-8041.
- Initiated
- 2013-05-01
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-02-06
- Distribution
- Worldwide Distribution: USA Nationwide and countries of: Argentina, Austria, Australia, Bahrain, Belgium, Canada, Chile, Columbia, Denmark, Ecuador, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Jordan, Kuwait, Malaysia, Lebanon, Lybia, Mexico, Netherlands, New Zealand, Pakistan, Peru, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Spain, South Africa, Switzerland, Taiwan-Province of China,Turkey, United Arabs Emirates United Kingdom, Uruguay, and Venezula-Bolivarian Republic.
- Recall number
- Z-1521-2013
- Recall status
- Terminated
- CDRH record ID
- 118293
- FDA classified
- 2013-06-07
- Recall event number
- 65217